Step 1: First you should hook the multi meter up to the test point zero (black lead) and to test point 3 (red lead). This allows you to read the voltage and to be able to set the solar cell and l.e.d. Lamp correctly.
Step 2: Next what you should do is adjust the solar cell and try to get the voltage to be set around negative four or as close as possible.
Step 3: Then after that is tightened down in place start adjusting the l.e.d. side with the lamp cover on. You want the window slits of lamp cover and output chute(which holds lenses) to match up. You also want to have the lamp to be touching the output chute.
Step 4: Once those are set in place remove the lamp cover and start adjusting the l.e.d.s with your finger to get the greatest negative voltage by moving them back and forth. Put the lamp cover back on but leaving it loose so you can adjust it to where you get the greatest negative voltage and tighten down. If your voltage is too high or low then what you do is take a regular screwdriver and turn the potentiometer. By turning the potentiometer clockwise this will lower your negative voltage and by turning it counterclockwise will give you a higher negative voltage.